Physical activity energy expenditure measured using a portable telemetric device in comparison with a mass spectrometer
1 Division of Human Nutrition, Faculty of Medicine and Surgery, University of Rome "Tor Vergata", Rome, Italy
2 Division of Radiology, Faculty of Medicine and Surgery, University of Rome "Tor Vergata"
Correspondence to:
Correspondence to:
Professor De Lorenzo, Division of Human Nutrition, Department of Neurosciences, Faculty of Medicine and Surgery, University of Rome "Tor Vergata", Via Montpellier, 1, 00133 Rome, Italy;
delorenzo{at}uniroma2.it
Background: Several studies have evaluated the accuracy of portable metabolic gas analysis systems, which measure physical activity energy expenditure. Since the Cosmed K2 telemetric system, two K4 devices (RQ and b2) have been developed.
Objective: To compare measurements using the Cosmed K4 RQ with measurements by a traditional method (mass spectrometer) for a sample of healthy subjects performing physical exercise of various intensities.
Methods: Nine healthy male footballers (mean (SD) age 18.3 (2.2) years) performed an incremental exercise running test from 8 km/h to exhaustion, with an increase of 2 km/h for each level of the test protocol. The expired air was collected simultaneously using the Cosmed K4 RQ and Airspec QP9000 mass spectrometer. Oxygen consumption (VO2) and carbon dioxide production (VCO2) were measured at rest and at each exercise intensity.
Results: No significant differences were found between VO2 and VCO2 measured by the two instruments, at rest or during exercise.
Conclusions: The application of the Cosmed K4 RQ system for VO2 and VCO2 measurements at rest and during both submaximal and maximal exercise may be of interest to sport physiologists, nutritionists, and doctors.
Keywords: energy expenditure; calorimeter; exercise; nutritional assessment
![]()
CiteULike
Complore
Connotea
Del.icio.us
Digg
Reddit
Technorati What's this?
This article has been cited by other articles:
-
Lepretre, P-M., Metayer, N., Giovagnoli, G., Pagliei, E., Barrey, E.
(2009). Comparison of analyses of respiratory gases made with the K4b2 portable and Quark laboratory analysers in horses. Vet Rec.
165: 22-25
[Abstract] [Full Text]
Register for free content
The full back archive is now available for all BMJ Journals. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006 right back to volume 1 issue 1. Register here to access the free archive of all BMJ Journals.
Don't forget to sign up for content alerts so you keep up to date with all the articles as they are published.