Register for email alerts and news feeds:
This journal | BMJ Group
rss
Published Online First: 3 September 2007. doi:10.1136/bjsm.2007.039552
British Journal of Sports Medicine 2008;42:212-216
Copyright © 2008 BMJ Publishing Group Ltd & British Association of Sport and Exercise Medicine.

ORIGINAL ARTICLES

Effects of hyperbaric exposures on cardiac pacemakers

V Lafay1, J A Trigano1, B Gardette2, C Micoli2 and F Carre3

1 CHP Résidence du Parc, Marseille, France
2 Comex, Marseille, France
3 CHU Pontchaillou, Rennes, France

Correspondence to:
Dr V Lafay, APHM, Marseille, France; vincent.lafay{at}medecins-saint-antoine.fr

Data on tolerance of cardiac pacemakers during diving are very scarce. The aim of this study was to test electronic and mechanical tolerances of pacemakers exposed to experimental reproductions of pressures encountered during diving. Two samples each of 20 different models of cardiac pacemakers were exposed to compression during continuous telemetric monitoring. The first sample of each model was exposed to a pressure of 60 metres of sea water (msw). Each second sample was first exposed to a pressure of 30 msw then to 60 msw hyperbaric testing, with a period of 1 month between the two tests. Electronic function and structural integrity of the cans were evaluated. No electronic dysfunction was noted. We merely observed in some devices a transient increase of the pacing rate during pressurisation. No significant deformation of the can (<=0.2 mm) was observed after the 30 msw hyperbaric test. However, after the 60 msw test, more than half of the devices tested were significantly and definitively deformed. These results show that tested pacemakers preserved a normal electronic function up to 60 msw but most of the tested devices demonstrated significant deformations of the pacemaker can for the hyperbaric exposure observed deeper than 30 msw. Without prejudging diving aptitude for implanted pacemaker patients, it therefore seems prudent to advise them against diving beyond 30 msw because of the potential for electronic dysfunction beyond that depth.


Add to CiteULike CiteULike   Add to Complore Complore   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us   Add to Digg Digg   Add to Reddit Reddit   Add to Technorati Technorati    What's this?

Relevant Article

Commentary on "Effects of hyperbaric exposures on cardiac pacemakers"
Benoìt Brouant
Br. J. Sports Med. 2008 42: 216. [Extract] [Full Text] [PDF]

This Article

Services
Citing Articles
Google Scholar
PubMed
Topic Collections
Bookmark with

Register for free content

The full back archive is now available for all BMJ Journals. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006 right back to volume 1 issue 1. Register here to access the free archive of all BMJ Journals.

Don't forget to sign up for content alerts so you keep up to date with all the articles as they are published.

 

The journal is co-owned by and the official journal of BASEM

Official journal of ECOSEP

Available online to all members of ACSP, AMSSM and SMNZ