Article Text
Abstract
Background: Several studies have evaluated the accuracy of portable metabolic gas analysis systems, which measure physical activity energy expenditure. Since the Cosmed K2 telemetric system, two K4 devices (RQ and b2) have been developed.
Objective: To compare measurements using the Cosmed K4 RQ with measurements by a traditional method (mass spectrometer) for a sample of healthy subjects performing physical exercise of various intensities.
Methods: Nine healthy male footballers (mean (SD) age 18.3 (2.2) years) performed an incremental exercise running test from 8 km/h to exhaustion, with an increase of 2 km/h for each level of the test protocol. The expired air was collected simultaneously using the Cosmed K4 RQ and Airspec QP9000 mass spectrometer. Oxygen consumption (Vo2) and carbon dioxide production (Vco2) were measured at rest and at each exercise intensity.
Results: No significant differences were found between Vo2 and Vco2 measured by the two instruments, at rest or during exercise.
Conclusions: The application of the Cosmed K4 RQ system for Vo2 and Vco2 measurements at rest and during both submaximal and maximal exercise may be of interest to sport physiologists, nutritionists, and doctors.
- energy expenditure
- calorimeter
- exercise
- nutritional assessment
Statistics from Altmetric.com
The metabolic cost of different physical activities, such as walking, running, working, and other sporting activities, in addition to the resting metabolic rate is useful for assessing the energy expenditure (EE) of a subject during a representative day.1 Whereas resting metabolic rate can usually be measured in clinical settings, EE during exercise is difficult to measure using indirect calorimetry.2
Over the last 30 years, technical innovations such as portable systems for metabolic gas analysis have allowed metabolic measurements of physical EE to be made in the office, field, and home settings. It has been suggested that these automated systems would increase the accuracy and reliability of exercise tests.3 However, it is important to check the accuracy and reliability of the flow and gas measurements. The Douglas bag technique, which is often used as reference method, has been shown to be valid and accurate,4 but the use of this system has some difficulties—for example, diffusion of gases through the bag.5,6 The Cosmed K2 portable device, which measures only oxygen uptake, has been shown to be accurate at rest7 and at both submaximal and maximal exercise levels.8,9 Successively, two Cosmed K4 devices (RQ and b2) containing a CO2 electrode have been developed. Only a few studies have examined the validity and reliability of these systems compared with a standard technique.6,10–12 We have previously reported an in vitro experimental study to validate the use of Cosmed K4 RQ (K4 RQ) for measuring flow rate and gas concentrations.13
The aim of this study was to compare in vivo measurements of the K4 RQ with measurements obtained using a traditional mass spectrometer for a sample of healthy subjects performing physical exercise of various intensities. We compare the accuracy and validity of Vo2 and Vco2 measurements between the two systems using simultaneous data collection.
METHODS
Subjects
Nine healthy male volunteer footballers (non-professional athletes) were selected randomly to participate in the study. Mean (SD) age, body weight, and height were 18.3 (2.2) years, 67.3 (4.6) kg, and 1.77 (0.57) m respectively. Each participant gave his written informed consent. The ethics committee of the University of Rome “Tor Vergata” approved the study protocol.
Instrumentation
The K4 RQ, as described previously,13 consists of three main components: a system to sample the expired air (mask and flowmeter); a portable transmitting unit (which measures 170 × 90 × 50 mm and weighs 550 g), fixed to a chest harness; and a receiving unit. It uses a Cosmed certified oxygen analyser (range 7–24%) and an infrared non-dispersive thermostatically controlled carbon dioxide analyser (range 0.0–8%). Before each test protocol, the instrument was calibrated against inspired air assumed to have a fractional oxygen concentration of 20.9% and a fractional carbon monoxide concentration of 0.03%, as described elsewhere.10,13 The flowmeter uses a bidirectional digital turbine and opto-electric reader, and, according to the manufacturer, has a linear response in the ventilation range 0.0–300 litres/min. The turbine flowmeter was calibrated using a 3 litre syringe at the beginning of the test protocol. This turbine was found to be accurate as a flowmeter.13,14 According to the technical description provided by the manufacturer, the 18 mm section turbine was used for measurements at rest, and the 28 mm section turbine was used for measurements during physical exercise. Oxygen consumption (Vo2) and carbon dioxide production (Vco2) were sampled every 15 seconds, and their values were measured at rest and at various exercise intensities (8, 10, 12, 14, 16, and 18 km/h).
The Airspec QP9000 (QP9000) is a quadrupole mass spectrometer housed in a robust high vacuum system. It is specifically designed for respiratory gas analysis and consists of a custom built dual microprocessor based controller (digital unit) connected to a personal computer. Levels of inspired and expired O2 and CO2 were determined at rest and at each level of the exercise protocol using this instrument.
Test protocol
The test protocol was conducted using the same tapis roulant (Technogym, Pesaro, Italy) for all participants, who performed an incremental exercise running test from 8 km/h to exhaustion, with an increase of 2 km/h for each level of the test protocol. The duration of each running test was set at six minutes. Only the mean values of measurements of the last two minutes were considered for further analysis. Before each exercise level, the participants rested for two minutes. The expired air was collected simultaneously for every subject using the K4 RQ and QP9000, which were disposed in parallel with a special small connector of low resistance respiratory value. Ventilation levels for Vo2 and Vco2 calculation were measured with the K4 RQ.
Data analysis
Statistical comparisons between Vo2 and Vco2 measured using the K4 RQ and QP9000 were made with an SPSS software package. Planned contrasts to limit comparisons between instruments for each condition were used. Mean values for measurements were analysed using Student’s paired t test. A significance level of p<0.05 was chosen for all comparisons.
RESULTS
Table 1 gives the mean values for Vo2, Vco2, and respiratory exchange ratio measured using the K4 RQ and QP9000 at rest and at different exercise levels (8, 10, 12, 14, 16, and 18 km/h). No significant differences were found between measurements using the two instruments either at rest or at any exercise level. Mean (SD) ventilation ranged from 8.8 (1.18) litres/min at rest to 116.50 (11.50) litres/min for the highest exercise level.
Oxygen consumption (Vo2, litres/min), carbon dioxide production (Vco2, litres/min), and respiratory exchange ratio (RER) measured with a Cosmed K4 RQ and an Airspec QP9000 mass spectrometer at rest and at different exercise levels (8, 10, 12, 14, 16, and 18 km/h) in nine male footballers
DISCUSSION
There are many reports of EE at rest and various workloads.3,5–17 Each athlete has unique energy requirements, which can be predicted by estimating resting metabolic rate and activity levels.18 Many studies have also evaluated the accuracy of portable metabolic gas analysis systems, which measure physical activity EE.6–14 An indirect calorimeter with a telemetric system is a convenient, accessible, and highly accurate instrument for measuring energy requirements of athletes. We have found that the portable metabolic system Cosmed K4 RQ is practical and precise for the determination of EE cost in athletes. Our study shows that data obtained with the K4 RQ portable calorimeter were not significantly different from those obtained with the Airspec QP9000 mass spectrometer. We compared Vo2 and Vco2 measurements using simultaneous data collection. McLaughlin et al11 found no significant difference between Vo2 at rest and at maximal workloads (250 W) measured using a K4 RQ and the Douglas bag method. However, the subjects were tested on consecutive days to avoid interference between the two methods, which may result in differences in Vo2 caused by day to day variation. One advantage of our study is that participants were tested on the same day. Vo2 and Vco2 measured with the K4 RQ were not significantly different from those measured with the QP9000 both at rest and at every exercise level.
The results confirm the validity of the K4 RQ system for gas measurements in the same subject at rest and during both submaximal and maximal exercise, suggesting its possible use for measuring EE over a wide range of exercise intensities. It may prove to be a valuable tool for optimising nutritional assessment of athletes. Measurement of EE during exercise may estimate the total EE compared with calculated predictions for nutritional needs. Application of the K4 RQ portable indirect calorimeter may be of interest to sport physiologists, nutritionists, and doctors.
Acknowledgments
We thank Dr Marcello Faina for invaluable help during the preparation of the article and Dr Ehab I Mohamed for discussion and help during the revision of the manuscript.