TY - JOUR T1 - Evaluation of the MyWellness Key accelerometer JF - British Journal of Sports Medicine JO - Br J Sports Med SP - 109 LP - 113 DO - 10.1136/bjsm.2009.062182 VL - 45 IS - 2 AU - S D Herrmann AU - T L Hart AU - C D Lee AU - B E Ainsworth Y1 - 2011/02/01 UR - http://bjsm.bmj.com/content/45/2/109.abstract N2 - Objective To examine the concurrent validity of the Technogym MyWellness Key accelerometer against objective and subjective physical activity (PA) measures. Design Randomised, cross-sectional design with two phases. The laboratory phase compared the MyWellness Key with the ActiGraph GT1M and the Yamax SW200 Digiwalker pedometer during graded treadmill walking, increasing speed each minute. The free-living phase compared the MyWellness Key with the ActiGraph, Digiwalker, Bouchard Activity cord (BAR) and Global Physical Activity Questionnaire (GPAQ) for seven continuous days. Data were analysed using Spearman rank-order correlation coefficients for all comparisons. Setting Laboratory and free-living phases. Participants Sixteen participants randomly stratified from 41 eligible respondents by sex (n=8 men; n=8 women) and PA levels (n=4 low, n=8 middle and n=4 high active). Results There was a strong association between the MyWellness Key and the ActiGraph accelerometer during controlled graded treadmill walking (r=0.91, p<0.01) and in free-living settings (r=0.73–0.76 for light to vigorous PA, respectively, p<0.01). No associations were observed between the MyWellness Key and the BAR and GPAQ (p>0.05). Conclusions The MyWellness Key has a high concurrent validity with the ActiGraph accelerometer to detect PA in both controlled laboratory and free-living settings. ER -